Transfer of the loaded, filtered active carbon without a
quartz container (vibration method, with polycarbonate
filters) in the quartz boat with holding-down clamp into
the multi-purpose combustion tube
EOX Injection of an extract via injection port with septum
(ABD
) into quartz glass boats and transfer into the multi
purpose combustion tube
EC/OC Transfer of the quartz-fiber filters or partial filters con-
taining deposits in the quartz glass boat with holding-
down clamp into the multi-purpose combustion tube
Sample volume
TS, TN, TX, TC,
TOC
Liquids 1 to 100 µl (horizontally with ABD)
1 to 500 µl (vertically with MMS or
TOC module
and direct manual dosing)
Solids 0.001 to 110 mg
Non-pressurized
gases
1 to 100 ml
Pressurized gases 1 to 20 ml (with
GSS/LPG combi module
)
1 to 100 ml (with GSS module and
GSS adapterbox
)
LPG 1 to 50 µl
EOX (extract) 10 to 100 µl
TOC (aqueous samples) 10 to 500 µl
Dosing speed (vertical)
TS, TN, TX, TC Liquids 0.2 to 2 µl/s
Recommended: 0,5 µl/s
Non-pressurized
gases
1 to 40 ml/min
Recommended: 20 ml/min
Pressurized gases Fixed (with GSS/LPG combi module)
1 to 40 ml/min
Recommended: 20 ml/min (with
GSS module
and GSS adapterbox)
LPG Solid
EOX 0.2 to 2 µl/s
Recommended: 0.5 µl/s
TC, NPOC 100 to 700 µl/s
Recommended: 350 µl/s
or manually adjusted
TIC Manually adjusted