Specifications multi EA 5000
178
Dosing speed (horizontal)
Liquids, EOX 1 to 10 µl/s
Recommended: 3 µl/s (with ABD +
MMS
or manual dosing)
ABD transfer speed to furnace controlled
automatically via flame sensor or soft-
ware settings
0,2 to 2 µl/s
Recommended: 0,5 µl/s (with
Autoinjector
)
Solids Fixed/transfer speed to furnace con-
trolled automatically via flame sensor or
software settings.
Non-pressurized gases 1 to 40 ml/min
Recommended: 20 ml/min
Pressurized gases Fixed (with GSS/LPG combi module)
1 to 40 ml/min
Recommended: 20 ml/min (with
GSS module
and GSS adapterbox)
LPG Solid
AOX Fixed/transfer speed to furnace con-
trolled via software settings for the ABD
Measuring gas dryer
TS, TN, TC, EC/OC Membrane dryer
TX, AOX, EOX Concentrated sulfuric acid
TOC, NPOC, TIC Condensation through Peltier cooling
Detection modules
Total nitrogen TN CLD 5000 Chemiluminescence
Total sulfur TS UVFD 5000 and
MPO 5000
UV fluorescence
Coulometric Sulfur Module Coulometric titration
AOX, EOX, TX, TCl, TOX, TIX Cl module Micro-coulometric end
point titration
(argentometry)
Total carbon TC, EC/OC TC module NDIR (non-dispersive
infrared absorption
spectrometry)
Total carbon TC, TIC, TOC,
NPOC, EC/OC
TOC module NDIR (non-dispersive
infrared absorption
spectrometry)
Sampling modules
Liquids Multi Matrix Sampler
MMS
Auto scaling
Autoinjector Semi-automatic
Solids ABD + MMS Auto scaling
Automatic Boat Drive
ABD
Semi-automatic
Non-pressurized gases GSS module

Table of Contents

Related product manuals