EasyManua.ls Logo

Endress+Hauser Analytik Jena multi EA 5000 - Page 179

Endress+Hauser Analytik Jena multi EA 5000
186 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
multi EA 5000 Specifications
179
Pressurized gases GSS/LPG combi module, GSS module + GSS adapterbox
LPG LPG module 2.0, GSS/LPG combi module
Process control
Control and analysis software multiWin
Software function scope (extract) Control of the analysis system, data log-
ging and analysis, recalculation, data ex-
port/import, maintenance wizard, on-
line help, real-time chart, service mod-
ule, report generator, automatic leak
check, self-check system
Gas supply
Oxygen Purity ≥ 4.5
Inlet pressure 600 kPa (6 bar)
Consumption Combustion 200 ml/min
Post-combustion 200 to 400 ml/min
Membrane dryer
drying flow
Approx. 500 ml/min
Argon Purity ≥ 4.6
Inlet pressure 600 kPa (6 bar)
Consumption Combustion 100 to 200 ml/min
Post-combustion 0 ml/min
Pneumatic seal
switch
Argon must be present
Argon protective
gas bypass (only
for the Cl module
)
Approx. 20 ml/min
Electrical variables
Voltage 110 to 240 V +10/-5 %
Frequency 50/60 Hz
Overvoltage category II
Fuses
Number of device fuses
T 10 AH
2
Typical average power consumption 1000 VA
PC interface 1 USB 2.0
Only use original fuses from Analytik Jena!
Environmental conditions
Temperature during operation +21 to 35 °C
Humidity during operation 90 % at 30 °C
Air pressure 0.7 to 1.06 bar
Temperature and humidity during stor-
age
+15 to 55 °C at 10 to 30 % humidity
(use desiccant)
Maximum elevation 2000 m

Table of Contents