multi EA 5000 Specifications
179
Pressurized gases GSS/LPG combi module, GSS module + GSS adapterbox
LPG LPG module 2.0, GSS/LPG combi module
Process control
Control and analysis software multiWin
Software function scope (extract) Control of the analysis system, data log-
ging and analysis, recalculation, data ex-
port/import, maintenance wizard, on-
line help, real-time chart, service mod-
ule, report generator, automatic leak
check, self-check system
Gas supply
Oxygen Purity ≥ 4.5
Inlet pressure 600 kPa (6 bar)
Consumption Combustion 200 ml/min
Post-combustion 200 to 400 ml/min
Membrane dryer
drying flow
Approx. 500 ml/min
Argon Purity ≥ 4.6
Inlet pressure 600 kPa (6 bar)
Consumption Combustion 100 to 200 ml/min
Post-combustion 0 ml/min
Pneumatic seal
switch
Argon must be present
Argon protective
gas bypass (only
for the Cl module
)
Approx. 20 ml/min
Electrical variables
Voltage 110 to 240 V +10/-5 %
Frequency 50/60 Hz
Overvoltage category II
Fuses
Number of device fuses
T 10 AH
2
Typical average power consumption 1000 VA
PC interface 1 USB 2.0
Only use original fuses from Analytik Jena!
Environmental conditions
Temperature during operation +21 to 35 °C
Humidity during operation 90 % at 30 °C
Air pressure 0.7 to 1.06 bar
Temperature and humidity during stor-
age
+15 to 55 °C at 10 to 30 % humidity
(use desiccant)
Maximum elevation 2000 m