Proline Prowirl F 200 PROFIBUS PA Diagnostics and troubleshooting
Endress+Hauser 143
Coding
(hex)
Measuring value
Limits
Quality SubstatusQuality
Byte 5
A0021271-EN
 32 Structure of the coding byte
The content of the coding byte depends on the configured failsafe mode in the particular
function block. Depending on which failsafe mode has been configured, status information
in accordance with PROFIBUS Profile Specification 3.02 is transmitted to the PROFIBUS
Master (Class 1) via the coding byte.
Determining the measured value status and device status via the diagnostic behavior
When the diagnostic behavior is assigned, this also changes the measured value status and
device status for the diagnostic information. The measured value status and device status
depend on the choice of diagnostic behavior and the group in which the diagnostic
information is located. The measured value status and device status are firmly assigned to
the particular diagnostic behavior and cannot be changed individually.
The diagnostic information is grouped as follows:
Diagnostic information pertaining to the sensor: diagnostic number 000 to 199
→  143
Diagnostic information pertaining to the electronics: diagnostic number 200 to 399
→  144
Diagnostic information pertaining to the configuration: diagnostic number 400 to 599
→  144
Diagnostic information pertaining to the process: diagnostic number 800 to 999
→  144
Depending on the group in which diagnostic information is located, the following
measured value status and device status are firmly assigned to the particular diagnostic
behavior:
Diagnostic information pertaining to the sensor (diagnostic no.: 000 to 199)
Diagnostic behavior
(configurable)
Measured value status (fixed assignment)
Device diagnostics
(fixed assignment)
Quality
Quality
Substatus
Coding
(hex)
Category
(NE107)
Alarm BAD
Maintenance
alarm
0x24 to 0x27
F
(Failure)
Maintenance
alarm
Warning GOOD
Maintenance
demanded
0xA8 to 0xAB
M
(Maintenance)
Maintenance
demanded
Logbook entry only
GOOD ok 0x80 to 0x8E
Off