Proline Prowirl F 200 PROFIBUS PA Diagnostics and troubleshooting
Endress+Hauser 159
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
485 Simulation measured variable Deactivate simulation Calculated saturated
steam pressure
Density
Energy flow
Flow velocity
Heat flow difference
Low flow cut off
Mass flow
Total mass flow
Switch output status
Pressure
Reynolds number
Specific volume
Corrected volume flow
Steam quality
Degrees of superheat
Temperature
Volume flow
Measured variable status
Quality Good
Quality substatus Function check
Coding (hex) 0xBC to 0xBF
Status signal C
Diagnostic behavior Warning
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
492 Simulation frequency output Deactivate simulation frequency output Calculated saturated
steam pressure
Energy flow
Flow velocity
Heat flow difference
Low flow cut off
Mass flow
Total mass flow
Switch output status
Corrected volume flow
Steam quality
Temperature
Volume flow
Measured variable status
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal C
Diagnostic behavior Warning
Diagnostic information Remedy instructions Influenced measured
variables
No. Short text
493 Simulation pulse output Deactivate simulation pulse output Calculated saturated
steam pressure
Energy flow
Flow velocity
Heat flow difference
Low flow cut off
Mass flow
Total mass flow
Switch output status
Corrected volume flow
Steam quality
Temperature
Volume flow
Measured variable status
Quality Good
Quality substatus Ok
Coding (hex) 0x80 to 0x83
Status signal C
Diagnostic behavior Warning