PROline Promag 10 11 Description of Device Functions
Endress+Hauser 101
ASSIGN STATUS
! Note!
This function is not available unless the STATUS setting was selected in the
OPERATION MODE function (Page 99).
Options:
ON (operation)
FAULT MESSAGE
NOTICE MESSAGE
FAULT & NOTICE MESSAGE
EPD (Empty Pipe Detection, only if active)
FLOW DIRECTION
VOLUME FLOW LIMIT VALUE
Factory setting:
FAULT MESSAGE
! Note!
The behaviour of the status output is a normally closed behaviour, in other
words the output is closed (transistor conductive) when normal, error-free
measuring is in progress.
It is very important to read and comply with the information on the switching
characteristics of the status output (see Page 103).
ON-VALUE
! Note!
This function is not available unless LIMIT VALUE or FLOW DIRECTION was
selected in the ASSIGN STATUS function.
Use this function to assign a value to the switch-on point (status output pulls
up). The value can be equal to, greater than or less than the switch-off point.
Positive and negative values are permitted.
User input:
5-digit floating-point number, [unit]
Factory setting:
0 [unit]
! Note!
The appropriate unit is taken from the SYSTEM UNITS group (see Page 93).
Only the switch-on point is available for flow direction output (no switch-off
point). If you enter a value not equal to the zero flow (e.g. 5 ), the difference
between the zero flow and the value entered corresponds to half the
switchover hysteresis.
OFF-VALUE
! Note!
This function is not available unless LIMIT VALUE was selected in the ASSIGN
STATUS function.
Use this function to assign a value to the switch-off point (status output drops
out). The value can be equal to, greater than or less than the switch-on point.
Positive and negative values are permitted.
User input:
5-digit floating-point number, [unit]
Factory setting:
0 [unit]
! Note!
The appropriate unit is taken from the SYSTEM UNITS group (see Page 93).
If SYMMETRY is selected in the MEASURING MODE function (Page 108)
and values with different signs are entered for the switch-on and switch-off
points, the notice message “INPUT RANGE EXCEEDED” appears.
Function description PULSE/STATUS OUTPUT