10 Technical data PROline Promag 10
80 Endress+Hauser
10.1.7 Operating conditions
Installation
Installation instructions Any orientation (vertical, horizontal)
Restrictions and additional installation instructions → Page 12 ff.
Inlet and outlet runs Inlet run: typically ≥ 5 x DN
Outlet run: typically ≥ 2 x DN
Length of connecting
cable
For the remote version the permissible cable length L
max
depends on the conductivity
of the medium → Page 21.
A minimum conductivity of 50 µS/cm is required for measuring demineralised water.
Environment
Ambient temperature
range
−20...+60 °C (sensor, transmitter)
Please note the following points:
• Install the device at a shady location. Avoid direct sunlight, particularly in warm
climatic regions.
• The transmitter must be mounted separate from the sensor if both the ambient and
fluid temperatures are high (→ “Fluid temperature”).
Storage temperature –10...+50 °C (preferably at +20 °C)
Degree of protection • Standard: IP 67 (NEMA 4X) for transmitter and sensor
• Optional: IP 68 (NEMA 6P) for remote version of Promag W and P sensor
Shock and vibration
resistance
Acceleration up to 2 g by analogy with IEC 68-2-6
CIP cleaning Promag W: not possible
Promag P: possible (note max. temperature)
Promag H: possible (note max. temperature)
SIP cleaning Promag W: not possible
Promag P: not possible
Promag H: possible (observe max. temperature)
Electromagnetic
compatibility (EMC)
To EN 61326
Emission: according to limit value for industrial environments documented in EN55011.