PROline Promag 10 11 Description of Device Functions
Endress+Hauser 111
11.11 Group SENSOR DATA
Function description SENSOR DATA
All sensor data (calibration factors, zero point and nominal diameter etc.) are set at the factory.
" Caution!
Under normal circumstances you should not change the following parameter settings, because changes
affect numerous functions of the entire measuring facility in general and the accuracy of the measuring
system in particular. For this reason, the functions described below are provided with an additional prompt
(with the code 10) once you enter your private code.
Contact the E+H service organisation if you have any questions about these functions.
K-FACTOR Use this function to display the current calibration factor (positive and negative
flow direction) for the sensor. The calibration factor is determined and set at
the factory.
User input:
5-digit fixed-point number: 0.5000 ...2.0000
Factory setting:
Depends on nominal diameter and calibration
! Note!
This value is also provided on the sensor nameplate.
ZERO POINT This function shows the current zero-point correction value for the sensor.
Zero-point correction is determined and set at the factory.
User input:
Max. 4-digit number: -1000...+1000
Factory setting:
Depends on nominal diameter and calibration
! Note!
This value is also provided on the sensor nameplate.
NOMINAL DIAMETER This function shows the nominal diameter for the sensor. The nominal diameter
depends on the size of the sensor and is set at the factory.
Options:
2...600 MM
1/12...24"
Factory setting:
Depends on the size of the sensor
! Note!
This value is also provided on the sensor nameplate.
MEASURING PERIOD Use this function to set the time for a full measuring period.
The duration of the measuring period is calculated from the rise time of the
magnetic field, the brief recovery time, the (automatically tracked) integration
time and the Empty Pipe Detection time.
User input:
40...1000 ms
Factory setting:
Depends on nominal diameter
EPD ELECTRODE Use this function to check whether the sensor is equipped with an EPD
electrode.
Display:
YES
NO
Factory setting:
YES Electrode fitted as standard