PROline Promag 10 3 Installation
Endress+Hauser 11
Transporting flanged devices (DN 350):
Use only the metal eyes on the flanges for transporting the device, lifting it and
positioning the sensor in the piping.
"
Caution!
Do not attempt to lift the sensor with the tines of a fork-lift truck beneath the metal casing.
This would buckle the casing and damage the internal magnetic coils.
Fig. 4: Transporting sensors with DN 350
3.1.3 Storage
Note the following points:
Pack the measuring device in such a way as to protect it reliably against impact for
storage (and transportation). The original packaging provides optimum protection.
The permissible storage temperature is 10...+50 °C (preferably +20 °C).
Do not remove the protective covers or caps on the process connections until you are
ready to install the device. This is particularly important in the case of sensors with
Teflon linings.
The measuring device must be protected against direct sunlight during storage in
order to avoid unacceptably high surface temperatures.
Choose a storage location where moisture does not collect in the measuring device.
This will help prevent fungus and bacteria infestation which can damage the liner.
F06-10Fxxxxx-22-xx-xx-xx-001