Technical data Proline Promag W 800
106 Endress+Hauser
Operating height According to EN 61010-1
• ≤ 2 000 m (6 562 ft)
• > 2 000 m (6 562 ft) with additional overvoltage protection (e.g. Endress+Hauser HAW
Series)
Atmosphere If a plastic transmitter housing is permanently exposed to certain steam and air mixtures,
this can damage the housing.
In cases of doubt, please contact the Sales Center.
Degree of protection Transmitter
• Standard: IP66/67, Type 4X enclosure, suitable for pollution degree 4
• When the housing is open: IP20, Type 1 enclosure, suitable for pollution degree 2
An inserted mating connector or a fitted protection cap is accepted as a prerequisite
for maintaining the specified degree of protection.
Proline Promag 800 sensor - Standard
• Standard: IP66/67, type 4X enclosure, suitable for pollution degree 4
• Optional or with order code "Calibration flow", option 8 "Water custody transfer approval":
• IP68 type 6P, suitable for pollution degree 4
• Order code "Sensor option", option CT "IP68, type 6P, 168h/3m (10 ft)" For use of
device under rain or surface water Use at a maximum water depth of 3 m (10 ft) for
168 h
Proline Promag 800 - Advanced sensor
• Standard: IP66/67, Type 4X enclosure, suitable for pollution degree 4
• Optionally available for remote version:
• IP66/67, type 4X enclosure; fully welded, with protective varnish EN ISO 12944 C5-M.
Suitable for use in corrosive atmospheres.
• IP68, type 6P enclosure; fully welded, with protective varnish as per EN ISO 12944 C5-
M. Suitable for permanent immersion in water ≤ 3 m (10 ft) or up to 48 hours at
depths ≤ 10 m (30 ft).
• IP68, type 6P enclosure; fully welded, with protective varnish as per EN ISO 12944
Im1/Im2/Im3. Suitable for permanent immersion in saline water ≤ 3 m (10 ft) or up to
48 hours at depths ≤ 10 m (30 ft) or in buried applications.
Accessories
Optional:
• External battery power supply: IP66/IP67, Type 4X enclosure
• Pressure measurement: IP68, 48 h under water 3 m (10 ft) with order code for
"Accessory enclosed", option PJ
• Pressure measurement: IP67 with order code for "Accessory enclosed", option PI
Vibration-resistance and
shock-resistance
Proline 800 - Advanced
Mount external mobile communications antenna remotely for any measuring points
that are subject to vibration.
Vibration sinusoidal, in accordance with IEC 60068-2-6
Compact version
• 2 to 8.4 Hz, 7.5 mm peak
• 8.4 to 2 000 Hz, 2 g peak
Remote version
• 2 to 8.4 Hz, 7.5 mm peak
• 8.4 to 2 000 Hz, 2 g peak
Vibration broad-band random, according to IEC 60068-2-64