Proline Promag W 800 Technical data
Endress+Hauser 105
Pressure
Pressure range, absolute [bar (psi)]
0.01 (0.1) ≤ p ≤ 8 (116)
8 (116) ≤ p ≤ 40 (580)
Measurement error, absolute
±0.5 % of 8 bar (116 psi)
±0.5 % o.r.
Accuracy of outputs
The outputs have the following base accuracy specifications.
Pulse output
o.r. = of reading
Accuracy Max. ±50 ppm o.r. (over the entire ambient temperature range)
Repeatability o.r. = of reading
Volume flow
Max. ±0.2 % o.r. ± 2 mm/s (0.08 in/s)
Electrical conductivity
Max. ±5 % o.r.
Influence of ambient
temperature
Pulse output
Temperature coefficient No additional effect. Included in accuracy.
16.7 Mounting
Mounting requirements →  22
16.8 Environment
Ambient temperature
range
→  28
Storage temperature The storage temperature corresponds to the operating temperature range of the
transmitter and the sensor →  28.
Protect the measuring device against direct sunlight during storage in order to avoid
unacceptably high surface temperatures.
Select a storage location where moisture cannot collect in the measuring device as
fungus or bacteria infestation can damage the liner.
If protection caps or protective covers are mounted these should never be removed
before installing the measuring device.
Relative humidity The device is suitable for use outdoors and indoors with a relative humidity of 4 to 95 %.