Proline Promag W 800 Mounting procedure
Endress+Hauser 43
PUSH TO
REMOVE
A0044279
Reverse the procedure to reassemble the measuring device.
6.2.6 Mounting the external battery pack
The external battery pack is mounted in the same way as the remote version of the
transmitter →  38.
6.3 Post-mounting check
Is the device undamaged (visual inspection)?
Does the measuring device conform to the measuring point specifications?
For example:
Process temperature →  107
Pressure (refer to the section on "Pressure-temperature ratings" in the "Technical Information"
document. →  120
Ambient temperature →  28
Measuring range →  96
Has the correct orientation been selected for the sensor →  25 ?
According to sensor type
According to medium temperature
According to medium properties (outgassing, with entrained solids)
Does the arrow on the sensor nameplate match the actual direction of flow of the fluid through
the piping →  25?
Are the measuring point identification and labeling correct (visual inspection)?
Is the device adequately protected against precipitation and direct sunlight?
Have the fixing screws been tightened with the correct tightening torque?