Test Setup - Test Configurator, Timer, and System
148 MAX-800 Series
RFC 2544 - Subtests
Measurement Mode, available with Dual Test Set, allows the selection
of the latency measurement mode for rates 10M to 10GE: Round-Trip
(default) or One-Way (MAX-880 only). For other rates, the
measurement mode is set to Round-Trip.
Synchronization with an external 1PPS clock is required to perform
One-Way Latency measurement. One-Way Latency is only possible
when both the local and remote 1PPS signal clocks are valid. The
following alarms are available with One-Way Latency measurement
mode.
LOPPS-L and LOPPS-R (Loss Of Pulse Per Second - Local/Remote) are
declared when either no pulse is received or when no pulse is
received within 1 second ± 6.6 μs after the previous pulse. LOPPS-R is
only monitored once the DTS connection is established.
Config. per Frame Size is available when the Copy From Throughput
check box is cleared and allows setting the Max. Rate for each frame
size. For Dual Test Set the Max. Rate is configurable for both local (L)
and remote (R) directions. For Dual Port topology, Max Rate is
configurable for both port directions.
All Frames check box when selected (cleared by default) allows
entering the maximum rate that will be applied to all frame sizes.