EasyManua.ls Logo

FEC Handheld Nutrunner HFC3000 - Page 57

FEC Handheld Nutrunner HFC3000
231 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4 Installation and Wiring
PAGE 4-21
4-6-1 How to Calibrate the External Monitoring Device
When pressing the CAL button on the unit display device, the torque voltage is output to the TORQUE
OUT pin with the voltage potential difference of approximately 3.7V (from tool ZERO level)
At the same time, the torque value detected on the tool is displayed on the DATA display screen.
Set the same torque value by using the external monitoring device to adjust the torque voltage.
(The zero torque voltage at no load is not 0V)
The voltage value of each tool varies. Therefore, the zero point needs to be corrected by using the
external monitoring device.
Output Circuit
Torque, Current, Speed Voltage Output Angle Pulse, Normal/Reverse Output
Current, Speed Voltage Output
The CAL voltage value is indicated at the
upper Display.
Press the CAL button.
Op amp output: 0 ~ ±5V
TTL (5V) output
+/- 25mA max
+/- 25mA max
Op amp output: 0 ~ ±10V

Table of Contents