FT-70 Fill Level Inspector Service Manual
5-40
X-ray Sample Cutoff Limit
This value is calculated automatically during the encoder
calibration. The normal value range is 3 to 4 and should be equal to
or 1 more than the number of X-ray Samples Per Encoder when
using a high speed production line.
X-ray Samples Per Encoder
This value verifies the correct X-ray Sample Cutoff Limit. It
displays the number of 500-microsecond x-ray samples that
occurred between the last two encoder interrupts. The encoder
interrupt is determined by the line speed and prescaler value.