EasyManua.ls Logo

Filtec FT-70c - Main Power System Troubleshooting

Filtec FT-70c
243 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FT-70 Fill Level Inspector Service Manual
8-16
Main Power System Troubleshooting
Power supply instability and failures account for many malfunc-
tions. We recommended testing the power supply voltages as a
starting point when troubleshooting. You need a digital voltmeter
with an impedance of at least 20,000 ohms per volt to perform
these procedures.
Testing the Power Supply
1. Remove the FT-70’s cover to access the power supply inside.
2. Using test point TP3 as ground, test all voltages at the appro-
priate test points.
Figure 8-1. Power Supply Circuit Board test points.
Reference
Test
Point
Output
VDC
Range
VDC
Xfmr
Winding
Source
Supply
Description
1 TP-6 +5 5.0/5.2 28 VAC +V Processor Logic, Scintillation Detector
2 TP-2 +12 11.5/12.5 28 VAC +V Scintillation Tube Pulse Preamplier,
Control Panel, Serial Port Drivers
3 TP-4 +22 18.0/22.0 28 VAC +V Sensor Emitters/Ampliers, Input
Comparators
4 TP-5 12 11.5/12.5 28 VAC Unreg Control Panel, Serial Port Drivers
5 J-1 +1000* 800/1200 28 VAC +22VDC Scintillation Detector
6 Q8-3 -5 4.9/5.1 28 VAC -12VDC RS232/RS422 Drivers
7 P4-2 +60 50/60 36 VAC Unreg Rejector Driver, Rejector
8 TP-3 Ground
1
4
2
3
8
5
ABCD
7
6
Industrial Dynamics Company, Ltd. © 40256-0600

Table of Contents

Related product manuals