Chapter 8: Diagnostic and Repair Procedures
8-13
Error Number 123
The X-ray Status Function is turned off. Turn the status back on to
activate x-ray generator.
Error Number 124
X-ray Source Disabled through external input connected to pin
TB4-5.
Error Number 125
X-ray Generator failure detected during warm-up cycle and the
generator will not activate. Check the LEDs on the generator.
If green LED is off the generator is not receiving power from
system.
Yellow Source Enabled LED should be on if unit is receiving
the signal to enable generation.
If red LED is off and the green LED is on, the source voltage is
low.
If red LED is on check the malfunction signal at pin TB4-5.
Error Number 126
X-ray generator is warming up for 3 minutes. This warm up
routine occurs if the generator is turned off for more than one
week. When the warm up s complete this error will automatically
clear.