FT-70 Fill Level Inspector Service Manual
8-20
Verify the Enable Signal at PX-1
1. Remove connector PX-1 and insert the tester point into pin 9.
2. Press Key 1 until X-RAY STATUS appears and press the Arrow
Key to turn the function on. If the tester indicates the signal is
received, the generator is malfunctioning and must be
replaced.
Verify the Enable Signal at TB4-13
1. Attach the tester point at TB4-13.
2. Press Key 1 until X-RAY STATUS appears and press the Arrow
Key to turn the function on. If the tester indicates the signal is
received, the signal cable from TB4-13 to PX-1 is faulty and
must be replaced.