Chapter 1: FT-70 System Overview
1-7
System Specifications
Specifications herein are listed for the FT-70. For specifications on
the rejector refer to the rejector manual that was included with
your FT-70.
Power Requirements
105 to 125 VAC, 50 or 60 Hz, 2.0 Amps, 1 Phase, or
220 to 250 VAC, 50 or 60 Hz, 1.0 Amp, 1 Phase
Input power must be computer grade and derived from a separate
branch circuit isolated from any equipment that can cause
electrical transients. The neutral line must be properly grounded.
Operating Temperature
Standard: 32-95 F (0-35 C)
Extended: 95-104 F (35-40 C)
Extreme (requires external cooling system): 104-122 F (40-50 C)
Pneumatics
The FT-70 is air purged to prevent contamination and corrosion
to the electronics housed inside. The air must be filtered, dry
compressed air at 80 to 120 psig with a volume of 0.5 CFM (138
to 827 kPa
@
14 liters/min).
X-Ray Generator
The X-ray Generator is a 30 KEV source that produces 45 KVP at 50
microamps.
Industrial Dynamics Company, Ltd. © Form 40256-0600