FT-70 Fill Level Inspector Service Manual
4-24
13. Press Key 2 until MINIMUM DIA MISSING LID appears and
then use the Arrow Key to enter the MD value you calculated.
14. Test the configuration by running several containers with and
without lids through the FT-70. If the containers are not
identified and rejected properly:
If sealed containers are identified as rejects, decrease the
Minimum Diameter Missing Lid value slightly and repeat
step 14 until false rejects no longer occur.
If unsealed containers are identified as good, increase the
Minimum Diameter Missing Lid value slightly and repeat
step 14 until rejects are no longer accepted.
15. If you are finished configuring the FT-70, press Key 5 until you
reach SYSTEM DISPLAY and then press the Arrow Key to
return to the Operate Mode.
GS BS()
2
-------------------------- B S+MD=
Industrial Dynamics Company, Ltd. © 40256-0600