Front Panel Operation
Setting the Output
4
4-31
12. Press the numeric keys and decimal point key to enter the desired frequency output
(maximum five numeric keys). Press a multiplier key, if necessary. For example,
press the kilo multiplier key K. Then press the H key. For example,
1.1234 kHz.
13. The Control Display now shows your voltage and frequency entries. For example,
123.456 mV and 234.567 mV at 1.1234 kHz (below).
1.1234 kHz
123.456 mV
234.567 mV
14. Press E. The calibrator clears your entry from the Control Display and copies
it into the Output Display (below is typical).
STBY
123.456 mV
234.567 mV
15. Press O to activate the calibrator output.
Two softkey labels appear on the Control Display: V@NOR/V@AUX and WAVE
MENUS.
WAVE
MENUS
PHASE
0.00
open
sine
sine
HARMONIC
MENUS
|LO|s
1123.4 Hz
WAVE
AUX WAVE
tied
tied
V ; NOR
V ; AUX
0 to
sine
tri
square
truncs
sine
tri
square
truncs
(180.00
V @ NOR (Voltage at NORMAL Terminals) V @ AUX (Voltage at AUX
Terminals) This is an information-only softkey position and does not have an
associated function. It shows the output function is dual ac voltage.