5500A
Operator Manual
1-24
1-27. Phase Specifications
1-Year Absolute Uncertainty, tcal ± 5 °C, (
∆Φ
Degrees)
10 to 65 Hz 65 to 500 Hz 500 Hz to 1 kHz 1 to 5 kHz 5 to 10 kHz
0.15 °
[1]
0.9 °
[2]
2.0 °
[3]
6 ° 10 °
[1] For 33 to 1000 V output, burden current <6 mA. For 6 to 20 mA burden current (33 to 330 V), the phase uncertainty is 0.4 degree.
[2] For 33 to 1000 V output, burden current <2 mA. For 2 to 5 mA burden current (33 to 330 V), the phase uncertainty is 1.5 degrees.
[3] For 33 to 1000 V output, burden current <2 mA. For 2 to 5 mA burden current (33 to 330 V), the phase uncertainty is 5 degrees.
Power Uncertainty Adder due to Phase Error ± (%)
Phase (Φ)
Watts
Degrees
Phase (Φ)
VARs
Degrees
PF
10 to 65 Hz 65 to 500 Hz 500 Hz to 1 kHz 1 to 5 kHz 5 to 10 kHz
0 90 1.000 0.00 0.01 0.06 0.55 1.52
5 85 0.996 0.02 0.15
10 80 0.985 0.05 0.29
15 75 0.966 0.07 0.43
20 70 0.940 0.10 0.58
25 65 0.906 0.12 0.74
30 60 0.866 0.15 0.92
35 55 0.819 0.18 1.11
40 50 0.766 0.22 1.33 Not Specified
45 45 0.707 0.26 1.58
50 40 0.643 0.31 1.88
55 35 0.574 0.37 2.26
60 30 0.500 0.45 2.73
65 25 0.423 0.56 3.38
70 20 0.342 0.72 4.33
75 15 0.259 0.98 5.87
80 10 0.174 1.49 8.92
85 5 0.087 2.99 17.97
90 0 0.000
To calculate exact ac Watts power adders due to phase uncertainty for values not shown, use the following formula:
()
()
Adder
Cos
Cos
%(
()
=−
+
100 1
Φ∆Φ
Φ
)
.
For example: for a PF of .9205 (Φ = 23) and a phase uncertainty of
∆Φ
= 0.15, the ac Watts power adder is:
()
()
Adder
Cos
Cos
%(
(23 )
=−
+
=100 1
15
23
0 11%
.
).
.