7-27. AC Voltage Harmonic Amplitude Accuracy (NORMAL)
Note
This verification test is optional. It is not necessary to guarantee the full
calibration of the instrument. However, it may be useful when
troubleshooting an instrument when other functions and/or parameters are
incorrect.
The AC Voltage Harmonic Amplitude Accuracy (NORMAL) tests the accuracy of the
harmonics from the NORMAL terminals. For this test, set the 5500A output to sinewave.
Nominal
Value (V)
(NORMAL)
Nominal
Value
(V)
(AUX)
Frequency
(Hz)
(AUX)
Harmonic
(NORMAL)
Frequency
(Hz)
(NORMAL)
Measured
Value (V)
(NORMAL)
Deviation
%
90-Day
Spec.
(%)
30 mV 300 mV 20 Hz 50th 1 kHz 0.243%
30 mV 300 mV 100 Hz 50th 5 kHz 0.243%
30 mV 300 mV 200 Hz 50th 10 kHz 0.243%
300 mV 300 mV 20 Hz 50th 1 kHz 0.053%
300 mV 300 mV 100 Hz 50th 5 kHz 0.053%
300 mV 300 mV 200 Hz 50th 10 kHz 0.053%
3 V 3 V 20 Hz 50th 1 kHz 0.024%
3 V 3 V 100 Hz 50th 5 kHz 0.024%
3 V 3 V 200 Hz 50th 10 kHz 0.024%
30 V 3 V 20 Hz 50th 1 kHz 0.034%
30 V 3 V 100 Hz 50th 5 kHz 0.034%
30 V 3 V 200 Hz 50th 10 kHz 0.034%
300 V 3 V 50 Hz 20th 1 kHz 0.044%
300 V 3 V 100 Hz 50th 5 kHz 0.070%
300 V 3 V 200 Hz 50th 10 kHz 0.070%
1000 V 3 V 50 Hz 20th 1 kHz 0.056%
1000 V 3 V 100 Hz 50th 5 kHz 0.170%
800 V 3 V 200 Hz 50th 10 kHz 0.275%
Optional:
1000 V
3 V 200 Hz 50th 10 kHz 0.250%