EasyManua.ls Logo

Fluke 5500A - DC Current Specifications

Fluke 5500A
410 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5500A
Operator Manual
1-12
1-17. DC Current Specifications
Absolute Uncertainty,
tcal ± 5 °C
± (% of output + µA)
Range
90 days 1 year
Resolution
Compliance
Voltage
Maximum
Inductive
Load
0 to 3.29999 mA 0.010 0.05 0.013 0.05
0.01 µA
4.5 V
1 µH
0 to 32.9999 mA 0.008 0.25 0.01 0.25
0.1 µA
4.5 V
200 µH
0 to 329.999 mA 0.008 3.3 0.01 3.3
1 µA
4.5 to 3.0 V
[1]
200 µH
0 to 2.19999 A 0.023 44 0.03 44
10 µA
4.5 to 3.4 V
[2]
200 µH
0 to 11 A 0.038 330 0.06 330
100 µA
4.3 to 2.5 V
[3]
200 µH
5725A Amplifier
0 to 11 A 0.03 330 0.04 330 100 4 V
400 µH
[1] The actual voltage compliance (Vc) is a function of current output (Io), and is given by the formula:
Vc = -5.05*Io+4.67. The highest compliance voltage is limited to 4.5 V.
[2] The actual voltage compliance (Vc) is a function of current output (Io), and is given by the formula:
Vc = -0.588*Io+4.69. The highest compliance voltage is limited to 4.5 V.
[3] The actual voltage compliance (Vc) is a function of current output (Io), and is given by the formula:
Vc = -0.204*Io+4.75. The highest compliance voltage is limited to 4.3 V.
Noise
Ranges
Bandwidth
0.1 to 10 Hz
p-p
Bandwidth
10 to 10 kHz
rms
0 to 3.29999 mA 20 nA 200 nA
0 to 32.9999 mA 200 nA
2.0 µA
0 to 329.999 mA 2000 nA
20 µA
0 to 2.19999 A
20 µA
1 mA
0 to 11 A
200 µA
10 mA
5725A Amplifier
0 to 11 A ± 25 ppm of output + 200 nA 2 mA

Table of Contents

Other manuals for Fluke 5500A