5500A
Operator Manual
1-28
1-33. AC Voltage (Non-Sine Wave) Specifications
1-Year Absolute Uncertainty,
tcal ± 5 °C,
± (% of output + % of range)
[2]
Triangle Wave &
Truncated Sine
Range
p-p
[1]
Frequency
% Output % Range
Maximum
Voltage Resolution
Normal Channel (Single Output Mode)
2.9 to 92.999 mV 0.01 to 10 Hz 5.0 0.5 Two digits on each range
93 to 929.999 mV 10 to 45 Hz 0.25 0.5
0.93 to 9.29999 V 45 Hz to 1 kHz 0.25 0.25
1 to 20 kHz 0.5 0.25
9.3 to 92.9999 V
20 to 100 kHz
[3]
5.0 0.5
Six digits on each range
Auxiliary Output (Dual Output Mode)
0.01 to 10 Hz 5.0 0.5 Two digits on each range
93 to 929.999 mV
10 to 45 Hz 0.25 0.5
45 Hz to 1 kHz 0.25 0.25
0.93 to 9.29999 V
1 to 10 kHz 5.0 0.5
Six digits on each range
[1] To convert p-p to rms for triangle wave, multiply the p-p value by 0.2886751. To convert p-p to rms for truncated sine wave,
multiply the p-p value by 0.2165063.
[2] Uncertainty is stated in p-p. Amplitude is verified using an rms-responding DMM.
[3] Uncertainty for truncated sine outputs is typical over this frequency band.
1-Year Absolute Uncertainty,
tcal ± 5 °C
± (% of output + % of range)
[2]
Square Wave
Range
p-p
[1]
Frequency
% Output % Range
Maximum
Voltage Resolution
Normal Channel (Single Output Mode)
2.9 to 65.999 mV 0.01 to 10 Hz 5.0 0.5 Two digits on each range
66 to 659.999 mV 10 to 45 Hz 0.25 0.5
0.66 to 6.59999 V 45 Hz to 1 kHz 0.25 0.25
1 to 20 kHz 0.5 0.25
6.6 to 65.9999 V
20 to 100 kHz 5.0 0.5
Six digits on each range
Auxiliary Output (Dual Output Mode)
0.01 to 10 Hz 5.0 0.5 Two digits on each range
66 to 659.999 mV
10 to 45 Hz 0.25 0.5
45 Hz to 1 kHz 0.25 0.25
0.66 to 6.59999 V
1 to 10 kHz 5.0 0.5
Six digits on each range
[1] To convert p-p to rms for square wave, multiply the p-p value by .5000000.
[2] Uncertainty is stated in p-p. Amplitude is verified using an rms-responding DMM.