5500A-SC600 Option
Oscilloscope Connections
8
8-13
8-17. Oscilloscope Connections
Using the cable supplied with the SC600 Option, connect the SCOPE output on the
Calibrator to one of the channel connectors on your oscilloscope (see Figure 8-1).
To use the external trigger, connect the TRIG OUT output on the Calibrator to the
external trigger connection on your oscilloscope. To use the external trigger and view its
signal with the calibration signal, connect the TRIG OUT output to another channel. See
your oscilloscope manual for details on connecting and viewing an external trigger.
5500A
CALIBRATOR
20V PK
MAX
HI
LO
TC
TRIG
OUT
1000V
RMS
MAX
20V
RMS
MAX
1V PK
MAX
20V PK
MAX
NORMAL
AUX
SCOPE
V, ,
RTD
A, -SENSE,
AUX V
200V PK
MAX
gl001f.eps
Figure 8-1. Oscilloscope Connection: Channel and External Trigger
8-18. Starting the SC600 Option
Press a (LED lit) to select the SC600 Option. The SCOPE menu, shown below,
appears in the Control Display. You can press any of the first four softkeys to go directly
to the VOLT, EDGE, LEVSINE, and MARKER calibration menus. Press the last softkey
to go to the OTHER menu (also shown below), allowing access to WAVEGEN, VIDEO,
PULSE, Impedance/Capacitance measurement (MEAS Z), and Overload (OVERLD)
menus. Press P to return to the SCOPE menu from the OTHER menu. This chapter
describes each of these menus in detail.
Select SCOPE mode:
VOLT EDGE LEVSINE MARKER OTHER
fffff
------------
|
---------------
|
------
Select SCOPE mode (or PREV MENU):
WAVEGEN VIDEO PULSE MEAS Z OVERLD
fffff