1-38. AC Current (Non-Sinewave) Specifications
Trianglewave &
Truncated Sinewave
1-Year Absolute Uncertainty,
tcal +
5°C,
+
(% of output + % of range)
[2]
Maximum
Current
Ranges
[1]
Frequency
%Output %Range Resolution
0.01 to 10 Hz 5.0 0.5 Two digits, e.g., 75 mA
10 to 45 Hz 0.25 0.5
45 to 1 kHz 0.25 0.25 Six digits on each range
2.9 to 92.999 mA
1 to 10 kHz 5.0 0.5
0.01 to 10 Hz 5.0 0.5 Two digits
10 to 45 Hz 0.25 0.5
45 to 1 kHz 0.25 0.5
93 to 929.999 mA
1 to 10 kHz 5.0 1.0
Six digits on each range
10 to 45 Hz 5.0 1.0 Two digits
45 to 1 kHz 0.5 0.5
0.93 to 6.19999 A
1 to 5 kHz 5.0 1.0
Six digits on each range
45 to 500 Hz 2.0 0.5 Two digits on each range
6.2 to 31 A
500 to 1 kHz 5.0 1.0 Six digits on each range
[1] All waveforms are peak-to-peak output ranges.
[2] Uncertainty is stated in peak-to-peak. Amplitude is verified using an rms-responding DMM.