EasyManua.ls Logo

Fluke 5500A - Leveled Sinewave Function Verification: Flatness

Fluke 5500A
410 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5500A
Operator Manual
8-92
Leveled Sinewave Function Verification: Amplitude (cont.)
Nominal Value (p-p) Frequency Measured Value (p-p) Deviation
(mV)
1-Year Spec.
(mV)
400.0 mV 50 kHz 8.200
500.0 mV 50 kHz 1.200
1.3 V 50 kHz 26.200
2.0 V 50 kHz 40.200
5.5 V 50 kHz 110.200
8-120. Leveled Sinewave Function Verification: Flatness
Nominal Value (p-p) Frequency Measured Value (p-p) Deviation (mV) 1-Year Spec. (mV)
5.0 mV 500 kHz 0.17
5.0 mV 1 MHz 0.17
5.0 mV 1 MHz 0.17
5.0 mV 2 MHz 0.17
5.0 mV 5 MHz 0.17
5.0 mV 10 MHz 0.17
5.0 mV 20 MHz 0.17
5.0 mV 50 MHz 0.17
5.0 mV 100 MHz 0.17
5.0 mV 125 MHz 0.20
5.0 mV 160 MHz 0.20
5.0 mV 200 MHz 0.20
5.0 mV 220 MHz 0.20
5.0 mV 235 MHz 0.20
5.0 mV 250 MHz 0.20
10.0 mV 500 kHz 0.25
10.0 mV 1 MHz 0.25
10.0 mV 1 MHz 0.25
10.0 mV 2 MHz 0.25
10.0 mV 5 MHz 0.25
10.0 mV 10 MHz 0.25
10.0 mV 20 MHz 0.25
10.0 mV 50 MHz 0.25
10.0 mV 100 MHz 0.25
10.0 mV 125 MHz 0.30