5500A
Operator Manual
7-14
7-14. AC Current Amplitude Accuracy
The AC Voltage Amplitude Accuracy test verifies the accuracy of ac current at the
5500A Calibrator front panel AUX terminals.
Nominal Value (A) Frequency (Hz) Measured Value (A) Deviation % 90-Day Spec. (%)
33 µA 1 kHz 0.848%
33 µA 10 kHz 1.395%
190 µA 45 Hz 0.169%
190 µA 1 kHz 0.222%
190 µA 10 kHz 1.019%
329 µA 10 Hz 0.236%
329 µA 45 Hz 0.136%
329 µA 1 kHz 0.166%
329 µA 5 kHz 0.346%
329 µA 10 kHz 0.986%
0.33 mA 1 kHz 0.171%
0.33 mA 5 kHz 0.241%
1.9 mA 1 kHz 0.096%
1.9 mA 10 kHz 0.466%
3.29 mA 10 Hz 0.159%
3.29 mA 45 Hz 0.089%
3.29 mA 1 kHz 0.089%
3.29 mA 5 kHz 0.159%
3.29 mA 10 kHz 0.459%
3.3 mA 1 kHz 0.161%
3.3 mA 5 kHz 0.241%
19 mA 1 kHz 0.086%
19 mA 10 kHz 0.466%
32.9 mA 10 Hz 0.159%
32.9 mA 45 Hz 0.079%
32.9 mA 1 kHz 0.079%
32.9 mA 5 kHz 0.159%
32.9 mA 10 kHz 0.459%
33 mA 1 kHz 0.161%
33 mA 5 kHz 0.241%
190 mA 1 kHz 0.086%
190 mA 10 kHz 0.466%
329 mA 10 Hz 0.159%
329 mA 45 Hz 0.080%
329 mA 1 kHz 0.080%