8-122. Marker Generator Function Verification
Nominal Interval Measured Interval Deviation 1-Year Spec.
4.98 s 25.12 ms
2.00 s 4.05 ms
999.21 ms 1.03 ms
500.00 ms 262.50 µs
200.00 ms 45.00 µs
100.00 ms 12.50 µs
50.00 ms 3.75 µs
20.00 ms 900.000 ηs
10.00 ms 350.00 ηs
5.00 ms 150.00 ηs
2.00 ms 54.000 ηs
1.00 ms 26.000 ηs
500.00 µs 12.750 ηs
200.00 µs 5.040 ηs
100.00 µs 2.510 ηs
50.00 µs 1.287 ηs
20.00 µs 0.506 ηs
10.00 µs 0.252 ηs
5.00 µs 0.125 ηs
2.00 µs 0.050 ηs
1.00 µs 0.025 ηs
500.000 ηs 0.013 ηs
200.000 ηs 5.000 ps
100.000 ηs 2.500 ps
50.000 ηs 1.250 ps
20.000 ηs 0.500 ps
10.000 ηs 0.250 ps
5.000 ηs 0.125 ps
2.000 ηs 0.050 ps