5500A
Operator Manual
8-30
8-43. Testing Video Triggers
Output at SCOPE
terminal (50).
LINE MK
10 ODD
FORMAT
ntsc
MODE
video
ff f f f
|
|
|
|
|
|
|
|
|
|
ntsc
pal
pal-m
secam
volt
edge
levsine
marker
wavegen
video
overld
meas Z
pulse
Mark = 10
New mark =
FIELD
odd
ff f f f
odd
even
You can press the MODE softkey to cycle through the functions in the order shown, or
you can press P to return directly to the OTHER modes menu.
Each option in the VIDEO menu is described below:
Output @ SCOPE terminal (50 ) Indicates the location of the signal output. If
the signal does not appear on the oscilloscope, press O. To disconnect the signal,
press Y.
LINE MK Allows you to select the marker line number. For ntsc and pal-m
formats, you can also select field (“odd” or “even”). For pal and secam formats, the
field (“ODD” or “EVEN”) is selected automatically based on marker line number.
FORMAT Scrolls through the available formats. You can select ntsc, pal, pal-m,
and secam.
MODE Indicates the calibrator is in VIDEO mode. Use the softkey to change
modes and open menus for other oscilloscope calibration modes.
Default video settings are + 100 %, format = NTSC, and videomark = 10.