5500A-SC600 Option
Testing the Trigger
8
8-29
The wave generator is available through the WAVEGEN menu, shown below. To access
this menu, press the softkey under MODE until “wavegen” appears.
Output @
SCOPE
WAVE
square
SCOPE Z
1 MΩ
OFFSET
+0.0V
MODE
wavegen
fff f f
square
sine
tri
1 MΩ
50Ω
volt
edge
levsine
marker
wavegen
video
pulse
meas Z
overld
You can press the MODE softkey to cycle through the functions in the order shown, or
you can press P to return directly to the OTHER modes menu.
Each option in the WAVEGEN menu is described below:
• OUTPUT @ SCOPE Indicates the location of the signal output. If the signal does
not appear on the oscilloscope, press O. To disconnect the signal, press Y.
• WAVE Scrolls through the three types of waveforms that are available. You can
select a square, sine, or triangle wave as the output.
• SCOPE Z Toggles the calibrator’s output impedance setting between 50 Ω and
1 MΩ.
• OFFSET Displays the offset of the generated wave. To change the offset, key in the
new value, and press E . Using the rotary knob does not change the offset; it
changes the actual voltage output.
When you change the offset, you must remain within certain limits to avoid clipping
the peaks. The limit depends on the wave’s peak-to-peak value. Specifically, the
peak excursion equals the absolute value of the offset plus half of the wave’s peak-
to-peak value. See “Wave Generator Specifications” at the beginning of this chapter.
• MODE Indicates you are in WAVEGEN mode. Use the softkey to change modes
and open menus for other oscilloscope calibration modes.
Default Wavegen settings are 20 mV p/p, 1000.0 Hz, WAVE = square, and
offset = 0.0 V.