8-86. Starting the Oscilloscope Calibration Option
Press a to start the Oscilloscope Calibration Option. The Control Display opens the
Volt menu, shown below, which contains options for calibrating the vertical gain on your
oscilloscope. This is the first of five calibration menus, which you can scroll through by
pressing the softkey under MODE. Each menu is described in detail in this chapter.
gl021i.eps
8-87. The Output Signal
The location of the output signal is indicated on the Control Display (the display on the
right side). If your 5520A is connected, but the output does not appear on the
oscilloscope, you may have the 5520A in standby mode.
The settings for the output signal are indicated in the Output Display (the display on the
left side). The following example shows the default settings for Volt mode, which are set
when you start the Oscilloscope Calibration Option.
gl022i.eps
If STBY is displayed, press the O key. The Output Display will show OPR and the
output should appear on the oscilloscope.