Fluke 5500A - Entering a Phase Angle

Fluke 5500A
410 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5500A
Operator Manual
4-44
The softkey PHASE is available after pressing the WAVE MENUS softkey that appears
when outputting dual ac voltages or ac power (shown below for ac power output).
I OUT WAVE
MENUSAUX
28.9589 mW
PHASE
0.00open
I WAVE
sine
V WAVE
sine
HARMONIC
MENUS
|LO|
1123.4 Hz
SHOW
PF
Phase = 0.00
New phase =
SHOW
PHASE
PF
lead
lead
lag
Power factor = 1.000
New pf =
When one output is a harmonic of the other, the phase shift is based on the phase angle
or power factor (cosine) of the harmonic signal. For example, when the AUX output is
generating a 60-Hz signal, and the NORMAL output is generating a 120-Hz (2nd
Harmonic) signal, a phase shift of 60° (pf of .5) would move the AUX signal 60° of
120-Hz (30° of 60 Hz).
4-41. Entering a Phase Angle
Complete the following procedure to enter a phase shift in degrees. This procedure
assumes you have already sourced a dual ac voltage or ac power output.
1. Press the softkey WAVE MENUS, opening the waveform menu.
2. Press the softkey PHASE, opening the phase entry menu.