5500A
Operator Manual
8-64
8-83. Trigger Signal Specifications for the Time Marker Function
Time Marker
Period
Division Ratio [1]
Amplitude into
50 (p-p)
Typical Rise
Time
5 to 1 s off/1 1 V 2 ns
0.5 to 0.1 s off/1/10 1 V 2 ns
50 ms to 100 ns off/1/10/100 1 V 2 ns
50 to 10 ns off/10/100 1 V 2 ns
5 to 2 ns off/100 1 V 2 ns
[1] Divider is internally limited to prevent trigger output from frequencies that are either
lower than 0.2 Hz (5 s period) or higher than 10 MHz.
8-84. Trigger Signal Specifications for the Edge Function
Edge Signal
Frequency
Division Ratio
Amplitude into
50 (p-p)
Typical Rise
Time
1 kHz to 1 MHz off/1 1 V 2 ns