7-12. AC Voltage Amplitude Accuracy (NORMAL)
The AC Voltage Amplitude Accuracy test verifies the accuracy of ac voltage at the
5500A Calibrator front panel NORMAL terminals.
Nominal Value (V) Frequency (Hz) Measured Value
(V)
Deviation % 90-Day Spec. (%)
30 mV 9.5 Hz 5.550%
30 mV 10 Hz 0.327%
30 mV 45 Hz 0.177%
30 mV 1 kHz 0.177%
30 mV 10 kHz 0.177%
30 mV 20 kHz 0.217%
30 mV 50 kHz 0.257%
30 mV 100 kHz 0.370%
30 mV 450 kHz 0.950%
300 mV 9.5 Hz 5.550%
300 mV 10 Hz 0.207%
300 mV 45 Hz 0.047%
300 mV 1 kHz 0.047%
300 mV 10 kHz 0.047%
300 mV 20 kHz 0.087%
300 mV 50 kHz 0.133%
300 mV 100 kHz 0.227%
300 mV 500 kHz 0.640%
3 V 9.5 Hz 5.550%
3 V 10 Hz 0.118%
3 V 45 Hz 0.022%
3 V 1 kHz 0.022%
3 V 10 kHz 0.022%
3 V 20 kHz 0.062%
3 V 50 kHz 0.110%
3 V 100 kHz 0.227%
3 V 450 kHz 0.490%
30 V 9.5 Hz 5.550%
30 V 10 Hz 0.118%
30 V 45 Hz 0.032%
30 V 1 kHz 0.032%
30 V 10 kHz 0.032%
30 V 20 kHz 0.069%
30 V 50 kHz 0.157%
30 V 90 kHz 0.227%