Fluke 5500A - AC Power Amplitude Accuracy (High Power)

Fluke 5500A
410 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5500A
Operator Manual
7-20
7-23. AC Power Amplitude Accuracy (High Power)
Note
This verification test is optional. It is not necessary to guarantee the full
calibration of the instrument. However, it may be useful when
troubleshooting an instrument when other functions and/or parameters are
incorrect.
The AC Power Amplitude Accuracy (High Power) tests the accuracy of the ac power
output at high power levels.
Nominal
Value (V)
(NORMAL)
Nominal
Value
(A)
(AUX)
Frequency
Phase
(degrees)
Measured
Value (V)
(NORMAL)
Deviation % 90-Day Spec.
329 V 2.19 A 5 kHz 0 0.065%
1 kV 11 A 1 kHz 0 0.048%