EasyManua.ls Logo

Fluke 5500A - DDS Assembly (A6)

Fluke 5500A
410 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Introduction and Specifications
Additional Specifications 1
1-31
1-39. AC Current (Non-Sine Wave) Specifications (cont)
1-Year Absolute Uncertainty,
tcal +
5°C,
+
(% of output + % of range)
[2]
Maximum
Current
Squarewave
Ranges
[1]
Frequency
%Output %Range Resolution
0.01 to 10 Hz 5.0 0.5 Two digits, e.g., 50 mA
10 to 45 Hz 0.25 0.5
45 to 1 kHz 0.25 0.25 Six digits on each range
2.9 to 65.999 mA
1 to 10 kHz 5.0 0.5
0.01 to 10 Hz 5.0 0.5 Two digits
10 to 45 Hz 0.25 0.5
45 to 1 kHz 0.25 0.5 Six digits on each range
66 to 659.999 mA
1 to 10 kHz 5.0 1.0
10 to 45 Hz 5.0 1.0 Two digits
45 to 1 kHz 0.5 0.5
0.66 to 4.39999 A
1 to 5 kHz 5.0 1.0
Six digits on each range
45 to 500 Hz 2.0 0.5 Two digits on each range
4.4 to 22 A
500 to 1 kHz 5.0 1.0 Six digits on each range
[1] All waveforms are peak-to-peak output ranges.
[2] Uncertainty is stated in peak-to-peak. Amplitude is verified using an rms-responding DMM.
1-40. AC Current, Square Wave Characteristics (typical)
Range Rise Time Settling Time Overshoot
I<4.4 A @ 400 Hz
25 µs 40 µs to 1 % of final value <10 % for loads <100
1-41. AC Current, Triangle Wave Characteristics (typical)
Linearity to 400 Hz Aberrations
0.3 % of p-p value, from 10 % to 90 % point <1 % of p-p value, with amplitude >50 % of range