Maintenance
Performing a Calibration Check
7
7-19
7-21. AC Power Amplitude Accuracy (High Voltage)
Note
This verification test is optional. It is not necessary to guarantee the full
calibration of the instrument. However, it may be useful when
troubleshooting an instrument when other functions and/or parameters are
incorrect.
The AC Power Amplitude Accuracy (High Voltage) tests the current outputs at the AUX
terminals in the presence of a high voltage.
Nominal
Value (V)
(NORMAL)
Nominal
Value
(A)
(AUX)
Frequency Hz
Phase
(degrees)
Measured
Value (A)
(AUX)
Deviation %
90-Day
Spec. (%)
1000 V 3.3 mA 65 Hz 0 0.161%
1000 V 3.3 mA 65 Hz 90 0.161%
1000 V 33 mA 500 Hz 0 0.161%
1000 V 33 mA 500 Hz 90 0.161%
1000 V 33 mA 1 kHz 0 0.161%
1000 V 33 mA 5 kHz 0 0.241%
1000 V 33 mA 7 kHz (10 kHz
optional)
0 0.541%
(Optional)
800 V
33 mA 10 kHz 0.541%
7-22. AC Power Amplitude Accuracy (High Current)
Note
This verification test is optional. It is not necessary to guarantee the full
calibration of the instrument. However, it may be useful when
troubleshooting an instrument when other functions and/or parameters are
incorrect.
The AC Power Amplitude Accuracy (High Current) tests the voltage outputs at the
NORMAL terminals in the presence of a high current.
Nominal
Value (V)
(NORMAL)
Nominal
Value
(A)
(AUX)
Frequency
Hz
Phase
(degrees)
Measured
Value (V)
(NORMAL)
Deviation % 90-Day Spec. (%)
33 mV 11 A 65 Hz 0 0.101%
33 mV 11 A 65 Hz 90 0.101%
330 mV 11 A 1 kHz 0 0.038%
3.3 V 2.19 A 5 kHz 0 0.048%
3.3 V 329 mA 10 kHz 0 0.048%