Fluke 5500A - Host Port Setup Procedure; Setting up the RS-232 Host Port for Remote Control

Fluke 5500A
410 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5500A
Operator Manual
5-10
5-5. Setting up the RS-232 Host Port for Remote Control
The 5500A Calibrator is fully programmable over an RS-232 link with a PC the rear
panel SERIAL 1 FROM HOST port (Figure 5-2). You can enter individual commands
from a terminal, write your own programs using, for example, a Windows-based
language such as Visual Basic, or run optional Windows-based Fluke software such as
5500/CAL or MET/CAL.
The RS-232 cable length for the port should not exceed 15 meters (50 feet), although
longer cable lengths are permitted if the load capacitance measured at a connection point
(including signal terminator) does not exceed 2500 pF.
5-6. RS-232 Host Port Setup Procedure
Complete the following procedure to set up the SERIAL 1 FROM HOST port. The
RS-232 parameters you select here must match the parameters set for the PC COM port.
The factory defaults (shown below in bold) are 9600 baud, 8 data bits, 1 stop bit, and no
parity. Other parameters include flow control, EOL (end-of-line) character, and EOF
(end-of-file) characters.
1. Turn the 5500A Calibrator power on. You may operate the calibrator during
warmup, but specifications are not guaranteed until warmup is complete.
2. Press the S key on the 5500A Calibrator front panel.
3. Negotiate the softkey selections shown below to select the serial port for remote
operation, then continue to Step 4.
Select
To step 4
gl43f.eps