Remote Operation
Setting up the IEEE-488 Port for Remote Control
5
5-9
5. At the prompt, type the following line to activate the IEEE interface card:
<ibdev 0 4 0 10 1 0>
The second number in this line is the primary address of the calibrator. If the address
has been changed from the factory default, change this line accordingly.
6. The prompt reads
<ud0:>. From this prompt type <ibwrt "remote"> then press the
ENTER (or RETURN) key.
7. Verify that the calibrator is now in remote control.
8. Select the Local command from the Control menu, then click OK in the Parameter
Input Window. Observe the Calibrator Control Display changes back to the reset
condition shown below.
nn323f.eps
9. From the ud0: prompt, type <q> and then press the ENTER (or RETURN) key.