8-65. Leveled Sinewave Verification: Flatness
Table 8-24. Leveled Sinewave Verification: Flatness
Nominal
Value (V p-p)
Frequency Measured
Value (V p-p)
Deviation
(V p-p)
1-Year Spec.
(V p-p)
0.005 50 kHz na na
0.005 30 MHz 0.000175
0.005 70 MHz 0.000175
0.005 120 MHz 0.0002
0.005 290 MHz 0.0002
0.005 360 MHz 0.0003
0.005 390 MHz 0.0003
0.005 400 MHz 0.0003
0.005 480 MHz 0.0003
0.005 570 MHz 0.0003
0.005 580 MHz 0.0003
0.005 590 MHz 0.0003
0.005 600 MHz 0.0003
0.0075 50 kHz na na
0.0075 30 MHz 0.0002125
0.0075 70 MHz 0.0002125
0.0075 120 MHz 0.00025
0.0075 290 MHz 0.00025
0.0075 360 MHz 0.0004
0.0075 390 MHz 0.0004
0.0075 400 MHz 0.0004
0.0075 480 MHz 0.0004
0.0075 570 MHz 0.0004
0.0075 580 MHz 0.0004
0.0075 590 MHz 0.0004
0.0075 600 MHz 0.0004
0.0099 50 kHz na na
0.0099 30 MHz 0.0002485
0.0099 70 MHz 0.0002485
0.0099 120 MHz 0.000298
0.0099 290 MHz 0.000298
0.0099 360 MHz 0.000496
0.0099 390 MHz 0.000496
0.0099 400 MHz 0.000496
0.0099 480 MHz 0.000496
0.0099 570 MHz 0.000496
0.0099 580 MHz 0.000496
0.0099 590 MHz 0.000496
0.0099 600 MHz 0.000496