Maintenance
Performing a Calibration Check
7
7-13
AC Voltage Amplitude Accuracy (NORMAL) (cont)
Nominal Value
(V)
Frequency (Hz) Measured Value
(V)
Deviation % 90-Day Spec.
(%)
300V 45 Hz 0.042%
300V 1 kHz 0.042%
300V 10 kHz 0.065%
300V 18 kHz 0.081%
1020V 45 Hz 0.048%
1020V 1 kHz 0.048%
1020V 5 kHz 0.160%
1020V 8 kHz (10 kHz optional) 0.200%
7-13. AC Voltage Amplitude Accuracy (AUX)
The AC Voltage Amplitude Accuracy test verifies the accuracy of ac voltage at the
5500A Calibrator front panel AUX terminals in the presence of a voltage at the
NORMAL terminals.
Nominal value
(V)
(NORMAL)
Nominal Value
(V)
(AUX)
Frequency
(Hz)
Measured
Value
(AUX V)
Deviation
%
90-Day Spec.
(%)
300 mV 10 mV 45 Hz 3.780%
300 mV 10 mV 1 kHz 3.780%
300 mV 10 mV 5 kHz 4.650%
300 mV 10 mV 10 kHz 4.800%
300 mV 300 mV 9.5 Hz 5.550%
300 mV 300 mV 10 Hz 0.273%
300 mV 300 mV 45 Hz 0.203%
300 mV 300 mV 1 kHz 0.203%
300 mV 300 mV 5 kHz 0.300%
300 mV 300 mV 10 kHz 0.450%
300 mV 3 V 9.5 Hz 5.550%
300 mV 3 V 10 Hz 0.165%
300 mV 3 V 45 Hz 0.085%
300 mV 3 V 1 kHz 0.085%
300 mV 3 V 5 kHz 0.197%
300 mV 3 V 10 kHz 0.347%
Note
The verification tests below are optional. It is not necessary to guarantee the full calibration
of the instrument. However, it may be useful when troubleshooting an instrument when other
functions and/or parameters are incorrect.
1000 V 10 mV 45 Hz 3.780%
1000 V 100 mV 1 kHz 0.450%
500 V 100 mV 5 kHz 0.600%
250 V 1 V 10 kHz 0.440%