Introduction and Specifications
General Specifications 1
1-13
1-18. Resistance Specifications
Absolute Uncertainty, tcal ± 5 °C
± (% of output + )
[2]
Range
[1]
90 days 1 year
Resolution
Allowable
Current
[4]
0 to 10.99
0.009 0.008
[3]
0.012 0.008
[3]
0.001 1 to 125 mA
11 to 32.999
0.009 0.015
[3]
0.012 0.015
[3]
0.001 1 to 125 mA
33 to 109.999
0.007 0.015
[3]
0.009 0.015
[3]
0.001 1 to 70 mA
110 to 329.999
0.007 0.015
[3]
0.009 0.015
[3]
0.001 1 to 40 mA
330 to 1.09999 k
0.007 0.06 0.009 0.06 0.01
250 µA to 18 mA
1.1 to 3.29999 k
0.007 0.06 0.009 0.06 0.01
250 µA to 5 mA
3.3 to 10.9999 k
0.007 0.6 0.009 0.6 0.1
25 µA to 1.8 mA
11 to 32.9999 k
0.007 0.6 0.009 0.6 0.1
25 µA to 0.5 mA
33 to 109.999 k
0.008 6 0.011 6 1
2.5 µA to 0.18 mA
110 to 329.999 k
0.009 6 0.012 6 1
2.5 µA to 0.05 mA
330 k to 1.09999 M
0.011 55 0.015 55 10 250 nA to 0.018 mA
1.1 to 3.29999 M
0.011 55 0.015 55 10
250 nA to 5 µA
3.3 to 10.9999 M
0.045 550 0.06 550 100
25 nA to 1.8 µA
11 to 32.9999 M
0.075 550 0.1 550 100
25 nA to 0.5 µA
33 to 109.999 M
0.4 5500 0.5 5500 1000
2.5 nA to 0.18 µA
110 to 330 M
0.4 16500 0.5 16500 1000
2.5 nA to 0.06 µA
[1] Continuously variable from 0 to 330 M.
[2] Applies for COMP OFF (to the 5500A Calibrator front panel NORMAL terminals) and 2-wire and 4-wire compensation.
[3] The floor adder is improved to 0.006 (0 to 10.99 range) and 0.010 (11 to 329.999 ) if the 5500A Calibrator is zeroed
(ohms zero or instrument zero) within 8 hours and temperature is ±1 °C of zeroing ambient temperature.
[4] Do not exceed the largest current for each range. For currents lower than shown, the floor adder increases by
Floor(new) = Floor(old) x Imin/Iactual. For example, a 100 µA stimulus measuring 100 has a floor uncertainty of 0.01 x 1
mA/100 µA = 0.1 .
Range
Maximum Voltage
[1]
Maximum Lead Resistance
[2]
0 to 10.99
1.37 V
<3.2
11 to 32.999
4.12 V
<3.2
33 to 109.999
7.7 V
<3.2
110 to 329.999
13.2 V
<3.2
330 to 1.09999 k
19.8 V
<6
1.1 to 3.29999 k
16.5 V
<6
3.3 to 10.9999 k
19.8 V
<6
11 to 32.9999 k
16.5 V
<6
33 to 109.999 k
19.8 V
<6
110 to 329.999 k
16.5 V
330 k to 1.09999 M
19.8 V
1.1 to 3.29999 M
16.5 V
3.3 to 10.9999 M
19.8 V
11 to 32.9999 M
16.5 V
33 to 109.999 M
19.8 V
110 to 330 M
19.8 V
(n/a 110 k and above)
[1] This is for the largest resistance for each range. The maximum voltage for other values is Imax (highest value of Allowable Current
above) multiplied by Rout.
[2] Maximum lead resistance for no additional error in 2-wire COMP.