1-19. AC Voltage (Sine Wave) Specifications
Absolute Uncertainty, tcal ± 5 °C
± (% of output + µV)
Range
Frequency
90 days 1 year
Resolution
Max
Burden
[1]
10 to 45 Hz 0.26 20 0.35 20
45 Hz to 10 kHz 0.11 20 0.15 20
10 to 20 kHz 0.15 20 0.2 20
20 to 50 kHz 0.19 20 0.25 20
50 to 100 kHz 0.26 33 0.35 33
1.0 to 32.999 mV
100 to 500 kHz 0.75 60 1 60
1 µV 50
10 to 45 Hz 0.19 50 0.25 50
45 Hz to 10 kHz 0.04 20 0.05 20
10 to 20 kHz 0.08 20 0.1 20
20 to 50 kHz 0.12 40 0.16 40
50 to 100 kHz 0.17 170 0.24 170
33 to 329.999 mV
100 to 500 kHz 0.53 330 0.7 330
1 µV 50
10 to 45 Hz 0.11 250 0.15 250
45 Hz to 10 kHz 0.02 60 0.03 60
10 to 20 kHz 0.06 60 0.08 60
20 to 50 kHz 0.10 300 0.14 300
50 to 100 kHz 0.17 1700 0.24 1700
0.33 to 3.29999 V
100 to 500 kHz 0.38 3300 0.5 3300
10 µV
10 mA
10 to 45 Hz 0.11 2500 0.15 2500
45 Hz to 10 kHz 0.03 600 0.04 600
10 to 20 kHz 0.06 2600 0.08 2600
20 to 50 kHz 0.14 5000 0.19 5000
3.3 to 32.9999 V
50 to 100 kHz 0.17 17000 0.24 17000
100 µV
10 mA
45 Hz to 1 kHz 0.04 6.6 mV 0.05 6.6 mV
1 to 10 kHz 0.06 15 0.08 15
33 to 329.999 V
10 to 20 kHz 0.07 33 0.09 33
1 mV
5 mA, except
20 mA for
45 to 65 Hz
45 Hz to 1 kHz 0.04 80 mV 0.05 80 mV
1 to 5 kHz 0.15 100 0.20 100
330 to 1020 V
5 to 10 kHz 0.15 500 0.20 500
10 mV
2 mA, except
6 mA for
45 to 65 Hz