Front Panel Operation
Sample Applications
4
4-57
12. Test the High current function.
a. Press R on the calibrator.
b. Verify that the calibrator is in standby and connect the DMM as shown in
Figure 4-17.
87
TRUE RMS MULTIMETER
MIN MAX
RANGE HOLD
H
REL Hz
mA
A
A
µ
mV
V
V
OFF
A
COM
V
mA
µ
A
PEAK MIN MAX
5500A
CALIBRATOR
20V PK
MAX
HI
LO
TC
TRIG
OUT
1000V
RMS
MAX
20V
RMS
MAX
1V PK
MAX
20V PK
MAX
NORMAL
AUX
SCOPE
V, ,
RTD
A, -SENSE,
AUX V
200V PK
MAX
f4-17.eps
Figure 4-17. Cable Connections for Testing an 80 Series High Amps Function
c. Set the calibrator output to 3.5 A at 0 Hz and press O. Verify the error is
within specification.
d. Repeat the previous step using 10.0 A at 0 Hz. Verify the error is within
specification.
e. Press Y on the calibrator and press the blue button on the DMM to switch to
ac measurements.
f. Set the calibrator output to 3.5 A at 60 Hz and press O. Verify the error is
within specification.
g. Repeat the previous step using the following calibrator settings:
AC Current Frequency
3.5 A 1.0 kHz
10.0 A 60 Hz
10.0 A 1.0 kHz
4-57. Calibrating the Meter
Continue with calibration if any range was out of tolerance in the previous procedure.
Note
The adjustment for calibrating the meter requires disassembling the meter.
Refer to the diagrams and access procedures in the 80 Series Service
Manual.