5500A-SC600 Option
SC600 Option Specifications
8
8-7
8-4. Edge Specifications
Table 8-2. Edge Specifications
Edge Characteristics into 50Load
1-Year Absolute
Uncertainty,
tcal ± 5 °C
Rise Time 300 ps (+0 ps / -100 ps)
Amplitude Range (p-p) 5.0 mV to 2.5 V ± (2% of output + 200 µV)
Resolution 4 digits
Adjustment Range ± 10% around each sequence value
(indicated below)
Sequence Values 5 mV, 10 mV, 25 mV, 50 mV, 60
mV, 80 mV, 100 mV, 200 mV,
250 mV, 300 mV, 500 mV, 600 mV,
1 V, 2.5 V
Frequency Range 1 kHz to 2 MHz [1] ± (2.5 ppm of setting)
Typical Jitter, edge to trigger < 5 ps (p-p)
Leading Edge Aberrations [2] within 2 ns from 50% of rising edge < (3% of output + 2 mV)
2 to 5 ns
< (2% of output + 2 mV)
5 to 15 ns < (1% of output + 2 mV)
after 15 ns < (0.5% of output + 2 mV)
Typical Duty Cycle 45% to 55%
Tunnel Diode Pulse Drive Square wave at 100 Hz to 100 kHz, with variable amplitude of 60 V to
100V p-p.
[1] The Frequency Range extends to 10 MHz with the Rise Time specification increased to 350 ps.
[2] All edge aberration measurements are made with a Tektronix 11801 mainframe with an SD26 input
module.