6. Remove the input signal by pressing Y.
7. Repeat this procedure for the remaining channels on your oscilloscope.
8-39. Calibrating the Time Base of an Oscilloscope
The horizontal deflection (time base) of an oscilloscope is calibrated using a method
similar to the vertical gain calibration. A time marker signal is generated from the
Calibrator and the signal’s peaks are matched to the graticule line divisions on the
oscilloscope.
8-40. The Time Marker Function
The Time MARKER function, which is available through the MARKER menu, lets you
calibrate the timing response of your oscilloscope. To access the MARKER menu, press
the softkey under MODE until “marker” appears.
Output at SCOPE
terminal (50).
SHAPE
spike
TRIG
off
MODE
marker
fffff
sine
spike
square
sq20%
off
/1
/10
/100
volt
edge
levsine
marker
wavegen
video
pulse
meas Z
overld
You can press the MODE softkey to cycle through the functions in the order shown, or
you can press P to return directly to the SCOPE menu.
Each option in the MARKER menu is described below:
OUTPUT @ SCOPE terminal (50 ) Indicates the location of the signal output. If
the signal does not appear on the oscilloscope, press O. To disconnect the signal,
press Y.