8-111. Verification Tables
Before the 5520A Scope Cal Option leaves the Fluke factory, it is verified to meet its
specifications at the following test points. The verification test points are provided here
as a guide when re-verification is desired.
8-112. Voltage Function Verification: AC Voltage into a 1 M
Ω
Load
Nominal Value (p-p) Frequency Measured Value (p-p) Deviation (mV) 1-Year Spec. (mV)
5.0 mV 10 Hz 0.11
5.0 mV 100 Hz 0.11
5.0 mV 1 kHz 0.11
5.0 mV 5 kHz 0.11
5.0 mV 10 kHz 0.11
10.0 mV 10 kHz 0.12
20.0 mV 100 Hz 0.15
20.0 mV 1 kHz 0.15
20.0 mV 10 kHz 0.15
50.0 mV 10 kHz 0.23
89.0 mV 10 Hz 0.32
89.0 mV 10 kHz 0.32
100.0 mV 10 kHz 0.35
200.0 mV 100 Hz 0.60
200.0 mV 1 kHz 0.60
200.0 mV 10 kHz 0.60
500.0 mV 10 kHz 1.35
890.0 mV 10 Hz 2.32
890.0 mV 10 kHz 2.32
1.0 V 100 Hz 2.60
1.0 V 1 kHz 2.60
1.0 V 10 kHz 2.60
2.0 V 10 kHz 5.10
5.0 V 10 Hz 12.60
5.0 V 10 kHz 12.60
10.0 V 10 kHz 25.10
20.0 V 10 kHz 50.10