Fluke 5500A - DC Power Amplitude Accuracy (NORMAL); DC Power Amplitude Accuracy (AUX)

Fluke 5500A
410 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5500A
Operator Manual
7-18
7-19. DC Power Amplitude Accuracy (NORMAL)
Note
This verification test is optional. It is not necessary to guarantee the full
calibration of the instrument. However, it may be useful when
troubleshooting an instrument when other functions and/or parameters are
incorrect.
The DC Power Amplitude Accuracy (NORMAL) tests the amplitude accuracy of the dc
power of the NORMAL terminals.
Nominal Value
(V)
(NORMAL)
Nominal Value
(A)
(AUX)
Measured Value
(V)
(NORMAL)
Deviation % 90-Day Spec. (%)
20 mV 2.19 A 0.020%
20 mV 11 A 0.020%
7-20. DC Power Amplitude Accuracy (AUX)
Note
This verification test is optional. It is not necessary to guarantee the full
calibration of the instrument. However, it may be useful when
troubleshooting an instrument when other functions and/or parameters are
incorrect.
The DC Power Amplitude Accuracy (AUX) tests the amplitude accuracy of the dc power
output of the AUX terminals.
Nominal Value
(V)
(NORMAL)
Nominal Value
(A)
(AUX)
Measured Value (A)
(AUX)
Deviation % 90-Day Spec. (%)
1000 V 100 µA 0.06%
1000 V 1 mA 0.015%
329 V 2.19 A 0.025%
1000 V 11 A 0.041%